Keep healing real with OCT
6 Aug 2009 by Evoluted New Media
If you want a high resolution, real-time instrument offering a non-contact, non-invasive technique with clinical applications - such as the dynamics of wound healing – you need OCT says Jon Holmes
If you want a high resolution, real-time instrument offering a non-contact, non-invasive technique with clinical applications - such as the dynamics of wound healing – you need OCT says Jon Holmes
“In clinical applications, OCT could function as an "optical biopsy" to image tissue microstructure in situ and in real-time.” |
The technique was conceived independently by Tanno and Chiba of Yamagata University in 1990, and by Huang and Fujimoto at MIT in 1991(1,2). The approach is analogous to ultrasound or Magnetic Resonance Imaging (MRI), except that imaging is performed by measuring the echo time delay and intensity of reflected laser light rather than sound or radio waves. Laser light is projected into tissue and, although most is 'scattered' and undetectable on its way back to the surface, a proportion of the light is not scattered (coherent - solid arrow) and is detected by an optical interferometer, which discards the scattered light (dotted arrows) and uses depth and intensity information to generate an image (Figure 1).
Although imaging depths are typically around 2mm, which is shallow compared with MRI or ultrasound, OCT can provide much higher image resolution on the micron scale. An optical focus provides lateral resolution and 2-D and 3-D cross-sectional OCT images of tissue are constructed by scanning the optical beam and performing axial measurements of light echoes at different transverse positions. However, until recently, it has not been possible to achieve high lateral resolution throughout the desired imaging depth. While axial resolution is determined by light source properties, the lateral resolution is a function of the numerical aperture (N.A.), which is a function of the desired depth of focus.
To overcome this fundamental limitation of traditional, single-beam FD-OCT
Figure 1: Light scattering paths in tissue |
Multi-Beam OCT technology has been implemented in the EX1301 OCT Microscope, suitable for ex-vivo applications, and in the new ‘VivoSight’ hand-held probe, designed for clinical applications and recently awarded a CE mark for clinical use.
To demonstrate the potential of Multi-Beam OCT for monitoring the progression of wound healing, an OCT image of skin on the back of the author’s hand was taken using a Michelson Diagnostics’ EX1301 OCT Microscope. This image is shown in Figures 2 and 3. The progress of the wound healing process is visible, including scab formation and gradual expulsion of the scab as the epidermis re-grows underneath.
OCT is already proving useful in ophthalmology and other clinical fields. Image
Figure 2: OCT Image of healthy skin immediately prior to infliction of small wounds. Image size 5mm X 2mm approx |
We have shown that Multi-Beam OCT may also be used to allow the non-
Figure 3: A series of five images taken over four days showing formation of scabs and re-growth of epidermis. Image size 5mm X 2mm approx |
Here we have shown that multi-beam OCT could be used for imaging applications in regenerative medicine, non-invasively, in real-time and at high-resolution. An instrument with clinical approvals will be available mid-2009.
For this and other clinical applications, the benefits offered by OCT are clear:
• Live sub-surface images at near microscopic resolution
• Instant, direct imaging of sub-surface microstructure
• No preparation of the sample required
• Eye-safe near infrared light used - no sample damage likely
• No ionising radiation - for use in office, clinic or lab
Figure 4: Tissue scaffold (image courtesy of University of Edinburgh) | Figure 5: Multi-Beam OCT image of a section of skin, approx 5mm by 1.4mm with a resolution <10μm |